• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Dexcom, Inc. Dexcom G7 Continuous Glucose Monitoring System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Dexcom, Inc. Dexcom G7 Continuous Glucose Monitoring System Back to Search Results
Catalog Number SW12300
Medical Device Problem Code Unintended Application Program Shut Down (4032)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Dexcom G7 Continuous Glucose Monitoring System
Common Device Name
Continuous Glucose Monitor
Manufacturer (Section D)
Dexcom, Inc.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
jaime chavez
6340 sequence dr.
san diego, CA 92121
+1(858)200-0200
MDR Report Key50954623
Report Number3004753838-2026-264636
Device Sequence Number10451861
Product Code QBJ
UDI-Device Identifier00386270001863
UDI-Public(01)00386270001863
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K213919
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberSW12300
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/17/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexFemale
Patient Weight230 LB
-
-