A device check on a patient's device revealed high impedance and a dc dc value of 4.The patient was reported to have an increase in seizures but it could not be determined if they were due to the high impedance, the age of the implanted generator or other factors.The patient was sent for x-rays which were not provided to the manufacturer for review.The physician elected to leave vns stimulation turned on since the patient was not experiencing any pain with stimulation and the patient was scheduled for revision surgery.During the revision surgery it was observed that the generator header had detached from the generator casing.The lead was disconnected from the header and the generator and lead were each explanted and replaced.Lead impedance with the new generator and lead was normal, 1402 ohms.No impedance testing of the explanted lead was performed.It is unknown if patient manipulation or trauma contributed to the header separation from the casing.It was reported that the patient is in a wheelchair and is lifted in and out of it frequently, however, the patient has not been involved in any accident or falls.It is unknown if the pre-surgical x-rays indicated evidence of the header separation.The physician was unable to determine the exact date of the increased seizures or if the seizure frequency was above or below the patient's pre-vns seizure rate.The patient's family stated, "it could have been more." the explanted generator and lead have not been received by the manufacturer for product analysis to date.
|
Based on the observed tool marks on the generator case and the condition of the header and casing in the detachment area, it is very likely that the reported header detachment occurred during the explant procedure.There was no evidence of dried body fluid or corrosion on the header or can in the detachment area.In order to test electrical function of the device a window was cut in the casing to facilitate output connections, however, an internal component was inadvertently damaged during the cutting of the window.After the component was replaced with a known good component, device testing was completed and the device was found to be operating properly according to all functional specifications.The lead assembly was returned in multiple portions, however, the electrodes and tie downs were not returned and therefore a complete evaluation could not be performed on the entire lead product.Abraded openings due to wear and incision marks were identified in multiple locations of the returned lead segments.Dried remnants of what appeared to have once been body fluids were observed inside the outer silicone tubing and likely entered via the cut ends of the lead segments.The condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.The setscrew marks found on the lead connector pins provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed during the visual analysis and no discontinuities were identified.Except for abraded openings in the outer tubing, there were no observed product related issues with the returned lead portions.Based on product analysis of the returned pulse generator and lead it is likely that the reported high lead impedance was due to a lead fracture in a segment of the lead which was not returned to the manufacturer for product analysis.
|