An explanted generator and lead were returned to the manufacturer for analysis in mfr.Report # 1644487-2015-05052.Product analysis was completed for the lead.No obvious anomalies were noted except for the set of setscrew marks found near the end of the connector pin, indicating the connector pin had not been fully inserted into the cavity of the pulse generator at one time.It appears that the incomplete pin insertion condition existed during the time of implant, as no other sets of setscrew marks were observed on the connector pin surface.There was no evidence to suggest an anomaly with the returned portion of the device.Note that since a portion of the lead assembly including the electrode array section was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Product analysis was completed for the generator.There were no performance or any other type of adverse conditions found with the pulse generator.It was found that impedance increased from 2268 ohms to 5717 ohms on approximately (b)(6) 2015.
|