Model Number SV-1010
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Low Blood Pressure/ Hypotension (1914); Hypoxia (1918); Itching Sensation (1943); Rash (2033)
Date of Event 09/18/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that immediately after the spaceoar vue placement procedure, the patient developed symptoms consistent with an anaphylactic reaction.The patient's spo2 dropped up to 80, the blood pressure was 50-60, itching and rash was observed on the upper extremities and trunk.The patient recovered after the administration of oxygen and intravenous fluids, solu-cortef (corticosteroid), and polaramine (antihistamine).The patient outcome was reported as stable.\ \ despite bsc's efforts, no further patient health information, underlying health concerns, details on the event and the patient's course after the event, or details regarding post procedure monitoring and observation have been received to date.If bsc receives additional pertinent information, it will be provided to the fda in a supplemental medical device report.
 
Additional Manufacturer Narrative
The device was not returned for analysis; therefore, complaint event could not be confirmed since there is no known device problem reported and the patient/impact codes reported are conditions that could not be tested.A dhr review was conducted and no deviations that could have contributed to the reported events were found.A labeling review confirmed the instructions for use (ifu) includes appropriate instructions for use, also the patient's symptoms are anticipated in the ifu.A risk review was completed and confirmed that the events of \"allergic reaction\" and \"rash, itching, hypotension and hypoxia\" (symptoms of allergic reaction as per medical review) were defined in the risk documentation.These events have been accounted during product risk analysis to support acceptable risk benefits for the product.Therefore, this complaint is assigned an investigation conclusion code of \"cause not established\" since the investigation findings do not lead to a clear conclusion.\ \ detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.\ \ block d4:h4\ the event was unable to provide the lot number of the device implanted.Therefore, there is not information related with device manufacturer day.\ \ block h6: \ imdrf patient code e1714 captures the reportable event of rash.\ imdrf patient code e1708 captures the reportable event of itching.\ imdrf patient code e2321 captures the reportable event of hypotension.\ imdrf patient code e0402 captures the reportable event of allergic reaction.\ imdrf patient code e0726 captures the reportable event of hypoxia.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SpaceOAR VUE System
Common Device Name
Absorbable perirectal spacer
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Cork Limited
500 commander shea boulevard
quincy MA 02171
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key51123771
Report Number2124215-2026-55072
Device Sequence Number10465409
Product Code OVB
UDI-Device Identifier00850009803009
UDI-Public(01)00850009803009
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K182971
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSV-1010
Device Catalogue NumberSV-1010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2026
Initial Report FDA Received Date10/16/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexMale
-
-