CYBERONICS, INC. PULSE GEN MODEL 102; GENERATOR Back to Search Results
Model Number 102
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Seizures (2063); Neck Pain (2433)
Event Type  Injury  
Event Description
It was reported that a patient's vns generator was explanted.The patient felt vns was making her seizures worse and was causing her pain in the neck.It is unknown if this pain was associated with stimulation.The patient also felt that vns did not help her seizures.Attempts for additional relevant information have been unsuccessful to date.The explanted generator has not been returned to the manufacturer for analysis to date.
 
Event Description
Further follow-up revealed that the physician indicated the increase in seizures was not related to vns.The seizures were not an increase above the patient's pre-vns baseline frequency.Clinic notes on (b)(6) 2015 note that the patient would like her vns removed because it is more trouble than it's worth.The explanted generator and lead were received for analysis.Analysis is underway, but has not been completed to date.
 
Event Description
Analysis of the generator was completed on 10/29/2015.The device performed according to functional specifications.Analysis in the pa lab concluded proper functionality of the pulse generator and that no abnormal performance or any other type of adverse condition.Analysis of the lead was completed on 11/06/2015.Note that since a portion of the lead (including the electrode array) was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Other than typical wear and explant related observations, no other anomalies were identified in the returned lead portion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 102
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5115756
MDR Text Key27202083
Report Number1644487-2015-05927
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Followup,Followup
Report Date 09/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/01/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date06/30/2012
Device Model Number102
Device Lot Number2740
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2015
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received10/29/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-