• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Nuvasive, Inc. ReLine; ORTHOSIS, SPINAL PEDICLE FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Nuvasive, Inc. ReLine; ORTHOSIS, SPINAL PEDICLE FIXATION Back to Search Results
Model Number 16027555
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 09/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experiencing back pain and imaging showed non-union loosening hardware.The revision surgery removed l4, l5, s1 screws and tlx20 cage and replaced with alif cage at l5-s1 and new screws.
 
Additional Manufacturer Narrative
The device has not yet been received for evaluation.The root cause cannot be determined at this time.A supplemental report will be submitted upon receipt of additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ReLine
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION
Manufacturer (Section D)
Nuvasive, Inc.
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
Globus Medical, Inc.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
+1(610)930-1800
MDR Report Key51187362
Report Number2031966-2026-00142
Device Sequence Number10447287
Product Code MNI
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K143684
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number16027555
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2026
Initial Report FDA Received Date10/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
Patient Weight156 LB
Patient RaceWhite
-
-