• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB TRULIGN TORIC POSTERIOR CHAMBER IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB TRULIGN TORIC POSTERIOR CHAMBER IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number BL1UT2700200
Device Problem Positioning Problem (3009)
Patient Problem Capsular Bag Tear (2639)
Event Date 09/14/2015
Event Type  Injury  
Manufacturer Narrative
Investigation is underway.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
It was reported that the lens was implanted after the ctr was put in place.The lens was rotated and the capsule tore.The lens would not center and vaulted.The lens was immediately removed.The surgeon believes that spinning the lens caused it to catch on the ctr which compromised the capsule.The lens was replaced with a 3-piece monofocal lens.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIGN TORIC POSTERIOR CHAMBER IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
BAUSCH + LOMB
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
10574 acacia street suite d1
rancho cucamonga CA 91730
Manufacturer Contact
sharon spencer
50 technology drive
irvine, CA 92618
9493985698
MDR Report Key5128991
MDR Text Key27519674
Report Number0001313525-2015-02680
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Report Date 09/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Model NumberBL1UT2700200
Device Lot Number7468016
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/14/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/29/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age47 YR
-
-