Model Number HAR36 |
Device Problems
Loss of Power (1475); Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/30/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
The complaint device was returned to stryker sustainability solutions for evaluation.Inspection of the returned device revealed evidence of clinical use including biological material on the distal tip and an indention in the teflon pad.The shaft rotation and the jaw actuation of the device were found to be acceptable.The device was connected to a generator and failed with two "tighten assembly" errors (after which the assembly was tightened) and a "replace instrument" error.The device was examined further and a build up of tissue/fluid was noticed along the rod which can disrupt harmonic frequency and reduce cutting ability.The rod was inspected for fractures, and none were found.The distal gasket on the rod was examined and revealed signs of damage and wearing of the gasket lip which would allow fluid/tissue to rise up the rod.The cause of tissue/fluid build up is damage to the distal gasket; action is being taken to address this issue.Tissue build up due to distal gasket damage can disrupt harmonic frequency and potentially reduce cutting ability; therefore, the reported complaint was confirmed.
|
|
Event Description
|
It was reported that har36 ultrasonic scalpel stopped working and then worked intermittently while grasping tissue during the case.There was no medical intervention, surgical delay, or adverse consequences reported.The procedure was completed successfully.
|
|