• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. OPT BLADELESS 12X100 STABILITY; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. OPT BLADELESS 12X100 STABILITY; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number 2B12LT
Device Problems Detachment Of Device Component (1104); Component Falling (1105)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/23/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Information anticipated, but unavailable at this time.When additional information is received and/or the device analysis has been completed, a supplemental medwatch will be sent.
 
Event Description
It was reported that during a laparoscopic sigmoidectomy, the duck-billed valve fell off when the obturator was inserted into the outer sleeve since the duck-billed valve was not set properly.The device was not used for the patient.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Batch # m91p9n.The analysis results found that the 2b12lt device was returned in good condition.During visual inspection, the duckbill was observed to be in good condition and properly seated; upon mechanical testing no issues were noted.The device was fully functional according to the manufacturing requirements.No conclusion could be reached as to what may have caused the reported incident.This report is not intended to deny that you experienced a problem with the device.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.The batch record was reviewed and no anomalies were noted during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPT BLADELESS 12X100 STABILITY
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 00969
Manufacturer Contact
milton garrett
4545 creek road ml 120a
cincinnati, OH 45242
5133378865
MDR Report Key5131956
MDR Text Key28066343
Report Number3005075853-2015-06302
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122512
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Followup
Report Date 09/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/18/2020
Device Catalogue Number2B12LT
Device Lot NumberM4HR0L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/18/2015
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-