Results: there was no visible damage to the exterior of the penumbra system aspiration pump max 110 (pump max).The pump was powered on and was confirmed to aspirate to specification of -25inhg.The vacuum gauge was tested to be within specification by using a calibrated vacuum gauge.Conclusions: evaluation of the returned device revealed that upon powering on, the pump was confirmed to aspirate to specification of -25inhg.In addition, the vacuum gauge was tested to be accurate using a calibrated vacuum gauge.The root cause of this complaint could not be determined.These devices are 100% functionally tested during inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110 (pump max).During the procedure, it was reported that the pump max produced adequate vacuum; however, the regulator knob (gauge) was not registering the pressure correctly, it read -18 inhg instead of -20 inhg.The procedure continued using the same pump max.There was no report of an adverse effect to the patient.
|