Model Number MODEL 250
Device Problems No Display/Image (1183); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/18/2015
Event Type  malfunction  
Event Description
It was reported that the medical professionals were having difficulties to use their programming system.It was reported that the handheld computer screen does not light up anymore.The physicians were having problems to adjust the handheld settings.
 
Event Description
The suspected handheld computer was returned to the manufacturer on 10/21/2015.Analysis is underway but it has not been completed to date.
 
Event Description
Analysis of the returned handheld computer was completed and no anomalies associated with the display were identified during the analysis.During the analysis, it was identified that the handheld would power down when the serial cable was inserted.The cause for the anomaly is associated with bent pins in the serial hood assembly that created an electrical short and caused the handheld to power down.Full handheld computer functionality was restored after replacement with a known good cable.No other anomalies were identified.Analysis of the software was completed and no anomalies associated with the flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5146710
MDR Text Key28657206
Report Number1644487-2015-06091
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1063524
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received11/18/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/13/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-