• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. LEAD MODEL 304

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. LEAD MODEL 304 Back to Search Results
Model Number 304-20
Device Problem High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/24/2015
Event Type  malfunction  
Event Description
It was reported that device diagnostics resulted in high impedance.The patient was referred to surgeon for generator and lead replacement.No known surgical interventions have occurred to date.
 
Event Description
The patient underwent generator and lead replacement due to high impedance.The generator was replaced prophylactically.There was no obviously lead break observed at the time of explant; however, the surgeon did not look for one.The lead was cut in multiple pieces during the explant process.The explanted lead and generator were received for analysis.Analysis is underway, but has not been completed to date.
 
Event Description
Analysis of the generator was completed on 12/01/2015.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.There were no performance or any other type of adverse conditions found with the pulse generator.Analysis of the lead was completed on 12/04/2015.A break was identified in the positive coil.Scanning electron microscopy images of the positive coil show that pitting or electro-etching conditions have occurred at the break location.However, due to metal dissolution, mechanical distortion (smoothed surfaces), and/or surface contamination.The fracture mechanism cannot be ascertained.Note that since the electrode array portion was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Other than the above mentioned observations and typical wear and explant related observations, no other anomalies were identified in the returned lead portions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 304
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5147745
MDR Text Key28574487
Report Number1644487-2015-06113
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 09/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/31/2017
Device Model Number304-20
Device Lot Number3518
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received12/01/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/09/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age35 YR
-
-