It was reported by the physician the patient had high impedance.It was later reported that patient had surgery on (b)(6) 2015 due to the high impedance observed.During surgery, the physician removed the vns generator from the pocket, removed the lead pin from the vns generator, wiped the lead pin off, re-inserted the lead pin into the vns generator, and tightened.After the lead pin was re-inserted into the vns generator, multiple diagnostics tests were run, both outside and inside the pocket, which showed the impedance was within the acceptable range.
|