Model Number CT7A |
Device Problem
Component Missing (2306)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional information become available, a supplemental record will be filed.
|
|
Event Description
|
The provider states the bolt in the frame by the fork is missing/broke off.
|
|
Manufacturer Narrative
|
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the headtube threads were stripped causing the threaded stud that attaches to the caster to come out, which confirmed the original complaint issue.
|
|
Event Description
|
The provider states the bolt in the frame by the fork is missing/broke off.
|
|