The generator is expected to be returned for analysis; however, analysis is not relevant to the wound dehiscence as the dehiscence was reported to be due to the patient's smoking.Product analysis, if the generator is received, will be reported in mfr.Report # 1644487-2015-05824.
|
The explanted lead was returned for analysis.It should be noted that only approximately 15cm of the lead was returned, not including the electrode array; therefore, an evaluation could not be performed on the entire lead.Setscrew marks were seen on the connector pin, providing evidence that, at one point in time, proper contact between the setscrew and the lead pin existed.Other than typical wear and explant related observations, no anomalies were identified in the returned portion of the lead.No other additional relevant information has been received to date.
|