Model Number 304-20
Device Problems High impedance (1291); Material Twisted/Bent (2981)
Patient Problem No Code Available (3191)
Event Date 01/01/2015
Event Type  malfunction  
Event Description
It was reported that the patient's device diagnostics showed high impedance with 9487 ohms.The patient underwent full revision surgery on (b)(6) 2015.Images of the old lead were provided from the surgery and the lead appeared to be twisted a lot.Patient manipulation is suspected as a result to be the cause of the lead fracture.The explanted devices were not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 304
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5225116
MDR Text Key31577218
Report Number1644487-2015-06427
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/31/2017
Device Model Number304-20
Device Lot Number202462
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received10/22/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/27/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age49 YR
-
-