Catalog Number 110940 |
Device Problems
Failure to Fire (2610); Temperature Problem (3022)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/15/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
|
Event Description
|
The surgeon was performing a partial knee arthroplasty using the robotic arm interactive orthopedic system (rio).During the case the anspach burr motor malfunctioned.The anspach motor was exchanged with a replacement anspach motor, which then also malfunctioned.The motor was exchanged for a third replacement motor.The case was then completed successfully.
|
|
Manufacturer Narrative
|
Reported event: anspach emax 2 plus burr motor completely inoperable.Method & results: -device evaluation and results: device evaluation was performed and the anspach emax 2 plus burr motor event was confirmed.-device history review: not performed as the anspach emax 2 plus burr motor is an oem product.-complaint history review: based on the device identification, the catsweb and trackwise complaint databases were reviewed from 2011 to present for similar reported events regarding anspach motor p/n 110940 s/n (b)(4) works intermittently.There have been no other events for the referenced serial number.Trend request for this part number has been submitted (trend request (b)(4)).Conclusions: the anspach emax 2 plus burr motor is an oem device.Upon receipt, the anspach emax 2 plus burr motor was bench evaluated by (b)(6) (engineer, rio robotics) and the reported event was confirmed.
|
|
Event Description
|
The surgeon was performing a partial knee arthroplasty using the robotic arm interactive orthopedic system (rio).During the case the anspach burr motor malfunctioned.The anspach motor was exchanged with a replacement anspach motor, which then also malfunctioned.The motor was exchanged for a third replacement motor.The case was then completed successfully.
|
|