Catalog Number 1070300-38
Device Problems Positioning Failure (1158); Material Rupture (1546); Unstable (1667)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/14/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that during a procedure of the heavily tortuous, mildly calcified, mid right coronary artery (rca), the 3.0 x 38 mm xience xpedition balloon was inflated to 18 atmosphere (atm) but ruptured during deployment and the partially expanded stent was not implanted.The stent was removed on the delivery system without reported issue and a second device was used to successfully complete the procedure.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the device was returned for analysis.The stent implant was stationary on the tightly folded balloon between the markers.There were stretched distal struts on the first two rings on the stent implant.There was no other damage noted to the stent implant.The stent implant was not loose as reported.The reported unstable stent, failure to deploy and material rupture were unable to be confirmed.Based on a visual, dimensional and functional inspection of the returned device, there is no indication of a product quality issue with respect to manufacture, design or labeling.Based on the information provided, the reported difficulties were unable to be confirmed; however, the noted damages appear to be related to circumstances of the procedure and not a product quality issue with respect to manufacture, design or labeling.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.Based on the information reviewed, there is no evidence to indicate the presence a potential quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key5258871
MDR Text Key32547003
Report Number2024168-2015-07184
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/13/2016
Device Catalogue Number1070300-38
Device Lot Number4100742
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/15/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-