Model Number GVTRSX58
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available a supplemental record will be filed.   (b)(6) 12/2/15 returns: the underlying cause documented on the irs or return fields in oracle is identified as: assembly/component issue/rear wheel cracked/broken.
 
Event Description
Provider alleges right rear wheel the bearing is rattling inside the wheel.
 
Manufacturer Narrative
Additional/updated information was added to reflect the device receiving an expanded evaluation.The result of the evaluation was that the frame was slightly bent causing the right front caster to three-wheel so that it only makes contact with the ground intermittently.When the wheels turn, a rattling and creaking noise can be heard coming from the right rear wheel bearings, confirming the complaint.However, the underlying cause could not be determined.
 
Event Description
Provider alleges right rear wheel the bearing is rattling inside the wheel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRSX5/WD08/AD/28/34P/U240/1255/U2222/AHDRM/9691B/GVT93HC/G1360/BH16 9153649403
Type of Device
WHEELCHAIR, MECHANICAL
Manufacturer (Section D)
INVAMEX
parque industrial manimex
reynosa 88780
MX  88780
Manufacturer (Section G)
INVAMEX
parque industrial manimex
reynosa 88780
MX   88780
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5290294
MDR Text Key33880321
Report Number9616091-2015-02863
Device Sequence Number1
Product Code IOR
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Replace
Type of Report Initial,Followup
Report Date 12/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberGVTRSX58
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/16/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-