• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 250
Device Problems Use of Incorrect Control/Treatment Settings (1126); Failure of Device to Self-Test (2937)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/15/2014
Event Type  malfunction  
Event Description
During review of the in-house programming/diagnostic history database, it was observed that on office visit on (b)(6) 2014 the patient's settings were different than what were programmed at the previous office visit on (b)(6) 2014.The settings found were indicative of a faulted diagnostic test; however, review of system diagnostic testing does not identify a faulted diagnostics that may have caused the change in settings.The settings were corrected on (b)(6) 2014.No patient adverse events were reported.
 
Event Description
It was reported that the patient was seen on (b)(6) 2014 and device settings upon interrogation were unchanged indicating that the fault did not occur until this visit.It was confirmed that this was the only visit prior to finding the change in settings which narrows the event down to this date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5294000
MDR Text Key34084493
Report Number1644487-2015-06729
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received01/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-