• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TIBIAL ALIGNMENT HANDLE; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TIBIAL ALIGNMENT HANDLE; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 6541-2-807
Device Problems Degraded (1153); Flaked (1246); Material Integrity Problem (2978)
Patient Problems No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
Event Date 11/20/2015
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
It has been reported by the customer that the handle of the product flakes little particles and falls into pieces.
 
Manufacturer Narrative
An event regarding santoprene handle degradation was reported.The event was confirmed.Method and results: device evaluation and results: visual analysis confirmed the reported event.The handle was returned degraded.Medical records received and evaluation: not performed because patient factors did not contribute to the reported event.Device history review: review of the device history records indicates that all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there has been one other event for the lot referenced.Conclusions: capa was initiated on 20-oct-2010 because a series of complaints were received for triathlon instrumentation where green santoprene over-mold handles have been reported to be degrading, becoming sticky and unstable.This event was determined to be under the scope of capa.Chemical and cytotoxicity tests showed the degraded santoprene is non-toxic.The ability to clean and sterilize the degraded instruments has not been compromised.
 
Event Description
It has been reported by the customer that the handle of the product flakes little particles and falls into pieces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ALIGNMENT HANDLE
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
beverly lima
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5312258
MDR Text Key34355662
Report Number0002249697-2015-04311
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeES
PMA/PMN Number
K123486
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/25/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2013
Device Catalogue Number6541-2-807
Device Lot NumberN1W04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received02/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/28/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-