STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Catalog Number 5532-G-509 |
Device Problems
Material Discolored (1170); Manufacturing, Packaging or Shipping Problem (2975); Packaging Problem (3007)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/07/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
When completed, the investigation results will be submitted in a supplemental report.
|
|
Event Description
|
The package had what looked like a dried liquid stain on the tyvek seal.
|
|
Manufacturer Narrative
|
An event regarding compromised packaging involving a triathlon ps x3 tibial insert was reported.The event was not confirmed.Method and results: device evaluation: the returned device confirmed the event as the (b)(6) of the outer blister had discoloration, along with the outer box.Clinician review: not performed as medical records were not received for review with a clinical consultant.Device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.-omplaint history review: there have been no similar previous reported events for this lot id.Conclusions: visual inspection of the returned device indicated a corner of the outer tyvek lid was discolored however, root cause could not be determined.The outer box showed signs of discoloration along one of the corners.Upon opening the box, it was noted that the stain on the outer blister aligned with the location of the stain on the outer box.The outer blister was discolored along the corner of the blister.The inner (b)(6) was unremarkable.
|
|
Event Description
|
The package had what looked like a dried liquid stain on the (b)(6) seal.
|
|