STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5532-G-509
Device Problems Material Discolored (1170); Manufacturing, Packaging or Shipping Problem (2975); Packaging Problem (3007)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2015
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
The package had what looked like a dried liquid stain on the tyvek seal.
 
Manufacturer Narrative
An event regarding compromised packaging involving a triathlon ps x3 tibial insert was reported.The event was not confirmed.Method and results: device evaluation: the returned device confirmed the event as the (b)(6) of the outer blister had discoloration, along with the outer box.Clinician review: not performed as medical records were not received for review with a clinical consultant.Device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.-omplaint history review: there have been no similar previous reported events for this lot id.Conclusions: visual inspection of the returned device indicated a corner of the outer tyvek lid was discolored however, root cause could not be determined.The outer box showed signs of discoloration along one of the corners.Upon opening the box, it was noted that the stain on the outer blister aligned with the location of the stain on the outer box.The outer blister was discolored along the corner of the blister.The inner (b)(6) was unremarkable.
 
Event Description
The package had what looked like a dried liquid stain on the (b)(6) seal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5315946
MDR Text Key34739571
Report Number0002249697-2015-04427
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2018
Device Catalogue Number5532-G-509
Device Lot NumberMMPK6H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received02/25/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-