• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE COYOTE¿; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE COYOTE¿; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number H74939185401510
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2015
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr.: the device was not returned for evaluation.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical/ procedural factors.(b)(4).
 
Event Description
It was reported that balloon rupture occurred.The 95% stenosed target lesion was located in the moderately tortuous and severely calcified superficial femoral artery (sfa).A 5f 45cm introducer sheath was placed.After a non-bsc guidewire was advanced to cross the lesion, a 4.0mm x 150mm x 150cm coyote balloon catheter was advanced for dilation.However, upon the first inflation, the balloon ruptured at 14 atmospheres.The device was completely removed and the procedure was completed with a different device.No patient complications were reported and the patient's status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE¿
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5317618
MDR Text Key34079069
Report Number2134265-2015-08940
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
PMA/PMN Number
K111295
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 12/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2018
Device Model NumberH74939185401510
Device Catalogue Number39185-40151
Device Lot Number18041150
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/03/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/31/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GUIDE WIRE: CRUISE
-
-