Model Number MODEL 250
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/02/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
A user reported that the tablet usb serial adapter cable would not attach properly to the programming wand.It was reported that the cable appeared to be bent due to apparent damage.To date, the usb serial adapter cable has not been returned to the manufacturer.
 
Event Description
The usb serial adapter cable was returned to the manufacturer for product analysis.The analysis determined that pins 4 and 5 of the db9 connector were bent, and pin 9 was missing.Once pins 4 and 5 were straightened, no anomalies associated with the cable performance were identified during the analysis.Device manufacturing records were reviewed and no unresolved nonconformances were noted prior to device distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5323377
MDR Text Key34926289
Report Number1644487-2015-06814
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 10.0
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 12/02/2015
Initial Date FDA Received12/23/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N