Results: the neuron max 6f 088 long sheath (neuron max) was kinked approximately 2.3 cm from the distal tip.The dilator was slightly bent near the distal tip.Conclusions: evaluation of the returned device confirmed that the neuron max was kinked near the distal tip.This damage likely occurred due to improper handling during removal from the packaging or during insertion into the patient.If the distal tip of the neuron max is pinched or manipulated at extreme angles, it is likely that the observed damage will occur.A kinked neuron max will likely contribute to the resistance indicated in the initial complaint.These devices are 100% visually evaluated during in-process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a medical procedure using a neuron max 6f 088 long sheath (neuron max).During the procedure, the physician had difficulty advancing the neuron max in the patient and it was removed.After removing the neuron max from the patient, the physician noticed that it was kinked near the dilator.The procedure was completed using another manufacturer's device.There was no report of an adverse effect to the patient.
|