• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
A follow-up report will be submitted once the investigation is complete.The patient height, and weight were not provided.
 
Event Description
The international customer reported the touch screen was temporarily unresponsive.There was no patient harm.
 
Manufacturer Narrative
The manufacturer's service technician was unable to duplicate the reported issue.The manufacturer's service technician recommended the touch screen be replaced to address the reported issue.The customer cancelled the repair, the device was returned to the customer without being repaired.The device successfully passed performance specification testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
robert corning
3000 minuteman road
andover, MA 01810
9093746996
MDR Report Key5332389
MDR Text Key35059363
Report Number2031642-2015-02781
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 12/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/09/2015
Initial Date FDA Received12/29/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/09/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age68 YR
-
-