Catalog Number 72200616S
Device Problem Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during an unknown procedure that the hand piece was overheating.
 
Manufacturer Narrative
Evaluation narrative - the subject device, one service replacement powermax elite, part number 72200616s was received on december 16th, 2015 and confirmed to be serial number (b)(4).Functional testing has found that the motor has seized, no additional test is possible with a seized motor.Unit was disassembled and inspected; corrosion was not found which would indicate a mechanical failure of the motor gearbox.The root cause of the overheated condition is most likely associated with the seized motor.A review of the device service records indicate that this serial number was delivered to the complainant in january of 2014.The complaint investigation has concluded that this unit has succumbed to expected wear and tear.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC REPL,MDU, HAND CNTRL, PWRMX
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
150 minuteman road
andover, MA 
5123585706
MDR Report Key5333180
MDR Text Key35037432
Report Number1643264-2015-00202
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/29/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/30/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse