MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CONQUEST 40 PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CONQUEST 40 PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number CQF7584
Device Problem Unraveled Material (1664)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2015
Event Type  malfunction  
Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that post procedure and upon removal from the introducer sheath, the pta balloon fibers were allegedly found to be unraveled.There was no reported impact or consequence to the patient.
 
Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Manufacturer Narrative
Manufacturing review: a manufacturing review was conducted.The lot met all release criteria.Visual/microscopic inspection: the device was returned.The balloon size for this product is printed on the balloon hub of the catheter and identified the returned sample as a 8mm x 4cm balloon.The balloon was examined and the fibers at the proximal end of the balloon had become unraveled but were still attached to the device.The unraveled fibers were located 4.5cm from the distal tip.The unraveled fibers extended from the balloon for approximately 19.0cm.No other anomalies were observed along the length of the device.Functional/performance evaluation: the patency of the guidewire lumen was tested using an in-house 0.035¿ guidewire, and it passed without issue.The inflation hub was connected to an inflation device and an attempt was made to inflate the balloon with water.While attempting to inflate the balloon to rbp, the balloon ruptured.This will be considered an incidental finding, as unraveled fibers likely contributed to the balloon rupturing during functional testing.The fibers were then stripped from the balloon and a longitudinal rupture was observed on the barrel of the balloon, beginning 4.1cm from the distal tip and continued proximally for 1.8cm.Medical records review: as medical records were not provided, a review could not be performed.Image/photo review: no medical images have been made available to the manufacturer.Conclusion: the investigation is confirmed for unraveled fibers on the barrel of the balloon.The definitive root cause could not be determined based upon the available information.It is unknown if patient and/or procedural issues contributed to the reported event.Labeling review: the current ifu (instructions for use) states: warnings:- do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.- when the catheter is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Precautions:- if resistance is felt during post procedure withdrawal of the catheter through the introducer sheath, determine if contrast is trapped in the balloon with fluoroscopy.If contrast is present, push the balloon out of the sheath and then completely evacuate the contrast before proceeding to withdraw the balloon.- if resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and guidewire/introducer sheath as a single unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST 40 PTA BALLOON DILATATION CATHETER
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 west 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V.
blvd montebello #1
parque industrial colonial
reynosa tamaulipas 88780
MX   88780
Manufacturer Contact
judith ludwig
1625 west 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key5336150
MDR Text Key35217915
Report Number2020394-2015-02075
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2018
Device Catalogue NumberCQF7584
Device Lot NumberREZH2142
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/17/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
-
-