STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PRIM CEM FXD BPLT #1; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5520B100
Device Problem Unstable (1667)
Patient Problem Synovitis (2094)
Event Date 09/29/2015
Event Type  Injury  
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.Not returned to the manufacturer.
 
Event Description
Patient underwent revision for mechanical problems.Op report notes "synovitis and instability." all components revised except patella component.
 
Manufacturer Narrative
An event regarding synovitis and instability involving a triathlon baseplate was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Medical records received and evaluation: not performed as medical records were not provided.Device history review: dhr review for the reported lot determined that the device was manufactured and packed to specification.Complaint history review: complaint history review confirmed that there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as product return, pre- and post-operative x-rays, operative reports as well as patient history and follow-up notes are needed to complete the investigation for determining a root cause.A capa trend analysis was conducted for the reported failure mode and concluded instability may result from other factors not necessarily related to the device.No further investigation is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Patient underwent revision for mechanical problems.Op report notes "synovitis and instability." all components revised except patella component.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PRIM CEM FXD BPLT #1
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
keyla navedo
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5339648
MDR Text Key34912946
Report Number0002249697-2016-00005
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2018
Device Catalogue Number5520B100
Device Lot NumberEEDAC
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/08/2015
Initial Date FDA Received01/01/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/03/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/24/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age63 YR
Patient Weight66
-
-