Catalog Number PXSLIM045-A
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/09/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2016-00012, 3005168196-2016-00013, 3005168196-2016-00015.The hospital disposed of the device.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the ovarian vein using ruby coils and px slim delivery microcatheters (px slim).During the procedure, the physician placed a px slim in the target vessel through another manufacturer's guide catheter.However, the physician was unable to advance a ruby coil through the px slim and they were both removed from the patient.The physician then attempted to advance two new px slim's through a guide catheter and another manufacturer's catheter but both px slim devices would not advance completely out of the catheters and were removed.The physician ended the procedure at this point.It is unknown if there was any adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PX SLIM DELIVERY MICROCATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key5342428
Report Number3005168196-2016-00014
Device Sequence Number10852234
Product Code DQY
UDI-Device Identifier00814548016405
UDI-Public00814548016405
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K100826
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Report Date (Section B) 12/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date03/31/2018
Device Catalogue NumberPXSLIM045-A
Device Lot NumberF62900
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date01/04/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-