• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #2 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #2 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5531G209
Device Problems Mechanical Problem (1384); Difficult To Position (1467); Mechanics Altered (2984); Positioning Problem (3009)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 12/10/2015
Event Type  malfunction  
Manufacturer Narrative
Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Two inserts could not be locked with the baseplate.Third insert could be locked.The baseplate was implanted.Two unlocked inserts could be obtained for the investigation.
 
Manufacturer Narrative
An event regarding a seating/locking issue involving a triathlon insert was reported.The event was confirmed.Method & results: device evaluation and results: visual inspection of the returned component indicates that the types of damage observed on the insert are indicative of an obstruction on the posterior aspect of the insert and on the retention slot of the inferior surface which may have been the reason that the user failed to position the insert correctly on the baseplate.Device history review: dhr review for the reported lot was satisfactory.Complaint history review: chr review for the reported lot confirmed that there are no other similar events reported.Conclusions: based on the visual inspection of the returned component, it appears that an obstruction on the posterior aspect of the insert and on the retention slot of the inferior surface may have been the reason that the user failed to position the insert correctly on the baseplate.A capa trend analysis was conducted for the reported failure mode and concluded that seating/locking difficulties may arise from user-related issues associated with the preparation of the joint space prior to the attempted seating and with the technique required to correctly introduce the insert component to the baseplate.No further investigation for this event is required at this time.
 
Event Description
Two inserts could not be locked with the baseplate.Third insert could be locked.The baseplate was implanted.Two unlocked inserts could be obtained for the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INSERT #2 9MM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5345260
MDR Text Key35385154
Report Number0002249697-2016-00014
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2020
Device Catalogue Number5531G209
Device Lot NumberLEP667
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2015
Initial Date FDA Received01/05/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/17/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/02/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age76 YR
Patient Weight60
-
-