One used device sample was returned for evaluation.Visual inspection revealed no abnormalities or product faults.The device was returned in the non-retracted position, with no site returned.Functional testing found the sample to operate as intended.When the device tab was pressed, the arm was found to easily pull into the retracted position, locking into place as intended.The reported product issue could not be duplicated during testing.Review of the device history records revealed no non-conformities during manufacturing.As the returned product was confirmed to operate as intended, no evidence was found to suggest the reported event was related to an intrinsic product problem.
|