ENCORE MEDICAL, L.P. MOVATION KNEE; MOVATION PS TIBIAL INSERT, SZ 6X9MM Back to Search Results
Catalog Number 292-09-006
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/09/2015
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to soft tissue.
 
Additional Manufacturer Narrative
Corrected data: model and lot #.The reason for this revision surgery was soft tissue.The length of in vivo service was 2.5 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records showed no non-conforming material reports associated with the main contributor component listed in the complaint.A review of the product complaint report history showed there have been 3 prior complaints reported against this part number; summary of investigations: 1 soft tissue, 1 instability, 1 infection.This is the first complaint against this lot number.This event is deemed to be non-product related.The root cause for this event and revision surgery is the result of soft tissue in the patient's knee.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOVATION KNEE
Common Device Name
MOVATION PS TIBIAL INSERT, SZ 6X9MM
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key5354106
Report Number1644408-2015-00774
Device Sequence Number13500428
Product Code JWH
UDI-Device Identifier00888912112321
UDI-Public(01)00888912112321
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K100900
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2015
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Followup
Report Date (Section B) 12/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/30/2017
Device Catalogue Number292-09-006
Device Lot Number141J1004
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/08/2016
Supplement Report FDA Received Date06/01/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient Weight109
-
-