• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM; MDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM; MDS Back to Search Results
Model Number 9500-35
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/15/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Patient's mother contacted dexcom on (b)(6) 2015 to report continuous glucose monitoring (cgm) inaccuracies compared to blood glucose (bg) meter that occurred on (b)(6) 2015.The sensor was inserted on (b)(6) 2015.It was reported the sensor was inserted in the arm.At the time of contact, no injury or medical intervention was reported.The device was not returned for evaluation.The receiver data log was provided by the patient.The data log was reviewed on 10/14/2015.The complaint of inaccurate cgm values was confirmed.It was also reported that the sensor was inserted in the arm.The dexcom g4® platinum (pediatric) continuous glucose monitoring system user's guide states: do not insert the sensor in sites other than the belly (abdomen) or upper buttocks.Other sites have not been studied and are not approved.Use in other sites might cause sensor glucose readings to be inaccurate and could result in you missing severe hypoglycemia (low blood glucose) or hyperglycemia (high blood glucose) events.However, a root cause for the reported inaccuracies cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
MDS
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key5358990
MDR Text Key35922240
Report Number3004753838-2016-30253
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 12/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/06/2016
Device Model Number9500-35
Device Catalogue NumberSTS-GL-007
Device Lot Number5201525
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/16/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/07/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age31 MO
Patient Weight16
-
-