| Model Number |
N/A |
| Medical Device Problem Codes |
Device Slipped (1584); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Pain (1994); No Code Available (3191); Subluxation (4525)
|
| Date of Event |
11/23/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.(b)(4).There are warnings in the package insert that state that this type of event can occur: under possible adverse effects, number 8 states, ¿dislocation and subluxation due to inadequate fixation and improper positioning.Muscle and fibrous tissue laxity can also contribute to these conditions.¿ this report is number 1 of 2 mdrs filed for the same event (reference 1825034-2016-00110 / 00111).
|
| |
|
Event or Problem Description
|
|
It was reported that patient underwent right total hip arthroplasty on (b)(6), 2015.Subsequently, patient underwent non-surgical repositioning on (b)(6), 2015 due to luxation.No further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.
|
| |
|
Event or Problem Description
|
|
There is no update to the prior event description provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4 g3 g6 h2 h6 mechanical: head h11.A review of complaint history identified additional similar complaints for the reported items and part and lot combinations.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.Medical records/radiographs were provided and reviewed by a health care professional.A review of the available records identified findings of the reported issues: findings: patient underwent a non-surgical repositioning performed requiring one day hospitalization and later had groin pain following repositioning.Musculoskeletal/connective tissue not captured as this is not an unexpected finding as the patient increases activity no change in root cause the event is confirmed via medical records.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
There is no update to the prior event description provided.
|
| |
|
Search Alerts/Recalls
|