This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.The instructions for use and patient manual include a reference guide for both visual and tone alarms including potential causes and actions to take.The patient manual outlines recommended practices for power management including expected battery behaviors and actions to take including replacement of devices which exhibit abnormal behavior.Additionally there is a warning to keep spare, fully charged batteries and back up controller available at all time.The steps and sequence for exchange of batteries and controllers are outlined.The instructions for use and patient manual outline the steps for handling and properly connecting power sources, including ensuring proper alignment, as well as instructions not to twist or force connections together in order to prevent damages.It is outlined to inspect the power connections and pins once a week, one at a time when changing the power source.A controller ac adapter and five batteries were returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the devices in relation to the reported event.Log file analysis revealed four critical battery alarms logged between (b)(6) 2015.The reported "double disconnect" could not be confirmed; no controller power up or motor start event was logged during the analyzed period.The returned battery passed external visual inspection and functional testing.Log file analysis reveal a critical battery alarm involving this battery; the battery was attached during the event but did not trigger the alarm.The battery was able to provide power and did not disconnect as alleged in the complaint details.Additionally, no bent pins were observed.The reported event could not be replicated or confirmed at the bench level; the battery perform as per specification.With review of the reported information and analysis of the returned device with no confirmed malfunction, a root cause cannot be determined.Heartware has opened an internal investigation to evaluate these types of issues.Heartware will submit a supplemental report when new facts arises which materially alters information submitted in a previous mdr report.This is five of six reports (3007042319-2015-01958, 3007042319-2015-01959, 3007042319-2015-01960, 3007042319-2016-00196, 3007042319-2016-00197 and 3007042319-2016-00198) submitted for devices related to the same event.
|