• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE FLORIDA OPERATIONS PLATINUM 10 CONCENTRATOR 9153642104; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE FLORIDA OPERATIONS PLATINUM 10 CONCENTRATOR 9153642104; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number IRC10LX
Device Problems Device Alarm System (1012); Loss of Power (1475); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available for the patient a supplemental record will be filed.
 
Event Description
Customer stated the concentrator was making a strange noise.Dealer picked up the unit and plugged it in.Unit shuts down without alarming.
 
Manufacturer Narrative
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the tie wraps on the product tank were leaking out o2 causing the noise, which confirmed the original complaint issue.However, there was no indication of a failure of the alarms.
 
Event Description
Customer stated the concentrator was making a strange noise.Dealer picked up the unit and plugged it in.Unit shuts down without alarming.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM 10 CONCENTRATOR 9153642104
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5366158
MDR Text Key35968958
Report Number1031452-2016-00149
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Repair
Type of Report Initial,Followup
Report Date 01/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberIRC10LX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/28/2015
Initial Date FDA Received01/14/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/10/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-