Model Number IRC10LX |
Device Problems
Device Alarm System (1012); Loss of Power (1475); Noise, Audible (3273)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional information become available for the patient a supplemental record will be filed.
|
|
Event Description
|
Customer stated the concentrator was making a strange noise.Dealer picked up the unit and plugged it in.Unit shuts down without alarming.
|
|
Manufacturer Narrative
|
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the tie wraps on the product tank were leaking out o2 causing the noise, which confirmed the original complaint issue.However, there was no indication of a failure of the alarms.
|
|
Event Description
|
Customer stated the concentrator was making a strange noise.Dealer picked up the unit and plugged it in.Unit shuts down without alarming.
|
|