MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. FOUNDATION KNEE; INSERT, STD., PRIMARY, SIZE10 9MM CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. FOUNDATION KNEE; INSERT, STD., PRIMARY, SIZE10 9MM CM Back to Search Results
Catalog Number 385-09-510
Health Effect - Clinical Code Pain (1994)
Date of Event 12/15/2015
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient presenting with pain when weight-bearing.The surgeon felt that with the age of the joint that the poly had worn and decided to replace it.
 
Additional Manufacturer Narrative
The reason for this revision surgery was pain due to the poly being worn.The length of in vivo service was 11.5 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records showed no non-conforming material reports associated with the main contributor component listed in the complaint.A review of the product complaint report history showed there have been 7 prior complaints reported against this part number; summary of investigations: 2 wear, 2 instability, 2 infections, 1 trauma.This is the first complaint against this lot number.This event is deemed to be non-product related.The surgeon believed the poly had worn as result of the implant age, reported no issues associated with the explanted product, and provided no information that definitively described the root cause or reason of the joint wear other than age of the insert.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION KNEE
Common Device Name
INSERT, STD., PRIMARY, SIZE10 9MM CM
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key5367418
Report Number1644408-2015-00792
Device Sequence Number518030
Product Code JWH
UDI-Device Identifier00888912117777
UDI-Public(01)00888912117777
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K000590
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2004
Device Explanted Year2015
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Followup
Report Date (Section B) 12/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2009
Device Catalogue Number385-09-510
Device Lot Number994811
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/14/2016
Supplement Report FDA Received Date06/09/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age63 YR
-
-