• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U128
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Patient Involvement (2645)
Event Date 10/06/2015
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).The product has been received for analysis.This report will be updated upon completion of analysis.
 
Event Description
Boston scientific received information that during pre-implant preparation, this cardiac resynchronization therapy pacemaker (crt-p) was unable to be interrogated with a programmer after multiple attempts.Another crt-p was able to be successfully interrogated and was used for implant.This device remained in the box and was never used.There was no patient involvement.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.The device had no telemetry and no pacing output.The pacemaker case was removed and the battery voltage was measured 0.890 volts, which, was noted to be too low to support device function.A thorough analysis of the battery was performed and laboratory analysis was unable to determine the cause of battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE
Type of Device
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5370707
MDR Text Key36234315
Report Number2124215-2015-13466
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 12/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/18/2016
Device Model NumberU128
Other Device ID NumberVALITUDE X4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received12/14/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-