Model Number K284
Device Problems Failure to Capture (1081); Low impedance (2285); Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/28/2015
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this pacemaker and right ventricular (rv) lead exhibited loss of capture and decreased impedances.It was reported that the patient was pacemaker dependent and experienced syncope while sitting in a chair.Subsequently, the patient underwent a revision procedure.Upon pocket opening it was discovered that the lead had not been fully advanced into the header of the pacemaker.The lead was successfully advanced and the setscrew was tightened.No further complications have been reported.This product remains in-service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGENIO
Type of Device
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5373301
MDR Text Key36230173
Report Number2124215-2015-16567
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/26/2017
Device Model NumberK284
Other Device ID NumberVITALIO
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received12/09/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/26/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age64 YR
-
-