• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; OZO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; OZO Back to Search Results
Model Number MMT-751NAS
Device Problem Thermal Decomposition of Device (1071)
Patient Problems Hypoglycemia (1912); Reaction (2414)
Event Date 01/04/2016
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Customer reported via phone call damage on the insulin pump, led anomaly and low blood glucose.Customer reported via phone call 49 mg/dl.Troubleshooting for low blood glucose was performed.Customer advised they went to the hospital as a result of low blood glucose levels and a severe reaction to insulin.Customer went through a hypoglycemic episode.Troubleshooting for cosmetic damage was performed.Customer advised the display screen was burnt and it was left out in the cold for 2 hours.Customer was advised to discontinue use of the device and revert to a backup plan.Customer was advised that the device would be replaced and agreed to return the product for analysis.
 
Manufacturer Narrative
A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing. after testing, it was concluded that the device operated within specifications. .
 
Manufacturer Narrative
Additional information has been provided by failure analysis that was not included on the initial device evaluation: the insulin pump passed the displacement accuracy test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP
Type of Device
OZO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key5377075
MDR Text Key36290472
Report Number3004209178-2016-82374
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
PMA/PMN Number
120010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup,Followup
Report Date 02/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-751NAS
Device Catalogue NumberMMT-751NAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received02/23/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age63 YR
Patient Weight77
-
-