Model Number MMT-751NAS |
Device Problem
Thermal Decomposition of Device (1071)
|
Patient Problems
Hypoglycemia (1912); Reaction (2414)
|
Event Date 01/04/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
Customer reported via phone call damage on the insulin pump, led anomaly and low blood glucose.Customer reported via phone call 49 mg/dl.Troubleshooting for low blood glucose was performed.Customer advised they went to the hospital as a result of low blood glucose levels and a severe reaction to insulin.Customer went through a hypoglycemic episode.Troubleshooting for cosmetic damage was performed.Customer advised the display screen was burnt and it was left out in the cold for 2 hours.Customer was advised to discontinue use of the device and revert to a backup plan.Customer was advised that the device would be replaced and agreed to return the product for analysis.
|
|
Manufacturer Narrative
|
A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing. after testing, it was concluded that the device operated within specifications. .
|
|
Manufacturer Narrative
|
Additional information has been provided by failure analysis that was not included on the initial device evaluation: the insulin pump passed the displacement accuracy test.
|
|