Catalog Number 9730492
Device Problems Unstable (1667); Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 12/21/2015
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Return requested.Replacement teratracker shipped to site 12/21/2015.Suspect teratracker discarded by site and will not be returned for analysis.No further issues have been reported.
 
Event Description
A medtronic representative reported a site's purple tera tracker screw was stripped.No further details regarding the damage, or how it occurred, were provided.There was no patient present when this issue was identified.
 
Manufacturer Narrative
Correction: contrary to what was previously reported the suspect teratracker was not discarded by site.Medtronic investigation of returned suspect device finds that the screw threads on the teratracker are slightly rounded but will thread into a known good tactile awl without issue.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERATRACKER, PURPLE
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
kristianna bilan
826 coal creek circle
louisville, CO 80027-9710
7208902338
MDR Report Key5377390
MDR Text Key36345789
Report Number1723170-2016-00068
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050651
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 05/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9730492
Device Lot Number150511
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received04/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-