Catalog Number 9730492 |
Device Problems
Unstable (1667); Material Deformation (2976)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/21/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.Return requested.Replacement teratracker shipped to site 12/21/2015.Suspect teratracker discarded by site and will not be returned for analysis.No further issues have been reported.
|
|
Event Description
|
A medtronic representative reported a site's purple tera tracker screw was stripped.No further details regarding the damage, or how it occurred, were provided.There was no patient present when this issue was identified.
|
|
Manufacturer Narrative
|
Correction: contrary to what was previously reported the suspect teratracker was not discarded by site.Medtronic investigation of returned suspect device finds that the screw threads on the teratracker are slightly rounded but will thread into a known good tactile awl without issue.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
|