• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 10° X3 INSERT 32MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 10° X3 INSERT 32MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 623-10-32D
Device Problems Device Dislodged or Dislocated (2923); No Apparent Adverse Event (3189)
Patient Problem Joint Dislocation (2374)
Event Date 12/22/2015
Event Type  Injury  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
Revision of a primary right hip due to dislocation.Surgeon revised the head and liner for a more stable construct.
 
Manufacturer Narrative
An event regarding an alleged dislocation involving a trident 10° x3 insert (liner) 32mm id was reported.Conclusions: review of the provided medical records and x-rays was performed by a clinical consultant concluded that the shell was placed in a horizontal position during primary surgery and that surgical technique contributed to dislocations and the need for a revision surgery in 2015.Additionally, the photos reveal impingement deformities, as seen on the inner lip of the liner.There was no indication that the trident 10° x3 insert contributed to this event.Based on the provided information, the product reported in this investigation did not contribute to the event.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Revision of a primary right hip due to dislocation.Surgeon revised the head and liner for a more stable construct.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 10° X3 INSERT 32MM ID
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5378355
MDR Text Key36343816
Report Number0002249697-2016-00137
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K033716
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2011
Device Catalogue Number623-10-32D
Device Lot NumberEVKMEA
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/25/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/24/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
Patient Weight90
-
-