An event regarding an alleged dislocation involving a trident 10° x3 insert (liner) 32mm id was reported.Conclusions: review of the provided medical records and x-rays was performed by a clinical consultant concluded that the shell was placed in a horizontal position during primary surgery and that surgical technique contributed to dislocations and the need for a revision surgery in 2015.Additionally, the photos reveal impingement deformities, as seen on the inner lip of the liner.There was no indication that the trident 10° x3 insert contributed to this event.Based on the provided information, the product reported in this investigation did not contribute to the event.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|