• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. RSP SHOULDER; SOCKET, INSERT 36MM SEMI RSP HUMERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. RSP SHOULDER; SOCKET, INSERT 36MM SEMI RSP HUMERAL Back to Search Results
Catalog Number 508-01-036
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 12/28/2015
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having an infection; and needing a poly and glenosphere swap.
 
Additional Manufacturer Narrative
The reason for this revision surgery was an infection.The previous surgery and the revision detailed in this investigation occurred 7 months apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) and product complaint report (pcr) database records show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Common Device Name
SOCKET, INSERT 36MM SEMI RSP HUMERAL
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key5381620
Report Number1644408-2016-00005
Device Sequence Number492992
Product Code KWS
UDI-Device Identifier00888912024709
UDI-Public(01)00888912024709
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K051075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2015
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Followup
Report Date (Section B) 12/28/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/30/2016
Device Catalogue Number508-01-036
Device Lot Number860C1010
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/20/2016
Supplement Report FDA Received Date04/26/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-