• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Catalog Number FEM14120
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/22/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history records are being reviewed.The event is currently under investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant was unable to provide any or procedural details to date.
 
Event or Problem Description
It was reported that the endovascular stent graft could not be deployed.Another stent graft was used to successfully complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.There is no indication for a manufacturing-related failure.The evaluation of the returned device confirmed the reported failure to deploy the stent graft.The stent graft was found to be partially released.The outer sheath was found to be elongated indicating that increased friction must have affected on the delivery system during the attempt to deploy the stent graft.During the performed function test, the stent graft could not be released.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.This type of event may be associated with a difficult vessel anatomy leading to increased friction and subsequent partial stent graft release.Insufficient flushing of the device may be another contributing factor to the reported event.On the basis of the information available, a definite root cause for the reported event could not be determined.The ifu states that the device must be flushed with sterile saline.Also the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure" and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
MDR Report Key5390134
Report Number9681442-2016-00023
Device Sequence Number223595
Product Code PFV
UDI-Device Identifier04049519008721
UDI-Public(01)04049519008721(17)180529(10)ANZE2714
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P130029/S003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Followup
Report Date (Section B) 12/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/29/2018
Device Catalogue NumberFEM14120
Device Lot NumberANZE2714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/26/2016
Supplement Report FDA Received Date04/07/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-