|
Results: the pet lock was intact on the proximal end of the penumbra coil 400 coil (pc400 coil) pusher assembly; the pusher assembly was kinked 24.0 cm from the proximal end and fractured approximately 25.0 cm; the proximal constraint sphere was intact with the distal detachment tip (ddt) with a piece of fractured stretch resistant wire (sr wire).The outer diameter (od) of the pusher assembly was measured and found to be within specification.Conclusions: evaluation of the returned device revealed the embolization coil's sr wire was fractured.This type of damage typically occurs due to improper handling during use.If the device was forcefully retracted against resistance, damage such as this may occur.Further evaluation revealed the pusher assembly was kinked and fractured.This type of damage typically occurs due to forceful advancement of the pc400 coil against resistance.The damage found on the pusher assembly and sr wire was not mentioned in the complaint.The od of the pusher assembly was measured and found to be within specification.The coil was not returned and, therefore, the od of the coil was not measured.Due to the embolization coil not being returned and the damage found on the pc400 coil, the root cause of the complaint could not be determined.Pc400 coils are 100% functionally tested during in-process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|