Model Number IRC5PO2V |
Device Problems
Device Emits Odor (1425); Smoking (1585)
|
Patient Problem
No Information (3190)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional information become available, a supplemental record will be filed.
|
|
Event Description
|
Dealer states the end user states it smells like something is burning inside, and the end user told the lincare driver he saw smoke, but the dealer could not duplicate.
|
|
Manufacturer Narrative
|
Product was returned for evaluation.The return fields in oracle state: stationary concentrators.Other, not able to duplicate issue.R: smells not duplicate e: alarming pilot valve not shifting e: leaking capacitor complaint was confirmed.The underlying cause could not be determined after reviewing the documentation in this investigation.
|
|
Event Description
|
Product was returned for evaluation.The return fields in oracle state: stationary concentrators.Other, not able to duplicate issue.R: smells not duplicate e: alarming pilot valve not shifting e: leaking capacitor dealer states the end user states it smells like something is burning inside, and the end user told the lincare driver he saw smoke, but the dealer could not duplicate.
|
|
Search Alerts/Recalls
|