|
The surgeon was performing a cervical fusion procedure at c6-c7.The surgeon followed the correct procedure by utilizing the punch awl before placing the screw.The punch awl angle trajectory and depth was analyzed under fluoro.A tcs swivel hex driver was used to insert a 3.5mm x 14mm locking screw into c6.During final x-ray, the cage and two screw construct were identified as being placed too off midline to the patient anatomy.The surgeon went to remove the screw in c6; however, the surgeon could not re-engage the hex driver into the screw head.Under the microscope, the surgeon identified the distal tip of the hex driver had partially broken off into the screw head.The implant was removed by the surgeon burring the bone of the c6 vertebrae directly above the screw hole.The removal of bone anterior to the screw allowed the surgeon to remove the implant device with the screw still engaged.The procedure was finished using another cervical standalone system.No adverse effects or injury to the patient as a result of the malfunction.
|