| Catalog Number |
48651234 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Injury (2348)
|
| Date of Event |
01/15/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient had an implant failure and subsidence of cervical grafts.The screws on the cervical plate backed out of the supposedly locked.The device was removed and replaced with non-stryker cervical plate.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: device history review; complaint history review; risk assessment.Results: the reported event was confirmed by the provided x-rays.Device was not returned.No relevant manufacturing issues were identified as all released units met stryker specifications.Conclusion: the plausible root cause of the reported event is not determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had an implant failure and subsidence of cervical grafts.The screws on the cervical plate backed out of the supposedly locked.The device was removed and replaced with non-stryker cervical plate.
|
| |
|