| Catalog Number |
1CX*FX15W |
| Medical Device Problem Code |
Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The actual device has not been returned to the manufacturing facility for evaluation.A follow up report will be submitted within 30 days of this report being sent.For this reason, evaluation code 11 was used in the conclusions section of h6.A review of the device history record and the product release decision control sheet of the involved product/lot # combination was conducted with no relevant findings.A search of the complaint file did not find any other report with the involved product code/lot# combination.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.9681834.Exemption number e2015022.All available information has been placed on file in quality assurance for appropriate tracking, trending and follow-up.H3 other text : device not returned to manufacturer.
|
| |
|
Event or Problem Description
|
|
The user facility reported that during prime the sampling line connection at the arterial line outlet was damaged on the capiox fx15 device.Follow up communication with the user facility confirmed the following information: (1) the oxygenator was changed out before the case; (2) surgery was completed successfully; and (3) there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 for mfg.Report no.9681834-2016-00004 to provide the retention sample evaluation results.The actual sample was not returned to the manufacturer for evaluation.Therefore, the investigation was based on evaluation of user facility information, and retention samples from product code 1cx*fx25w lot number 150615.Visual evaluation revealed no defects.The package configuration revealed that the joint between the blood out let port and the sampling line tube does not come into contact with any package materials.The retention sample was verified to be the normal product with no anomaly on the sampling tube.There is no evidence that this event was related to a device defect or malfunction and the exact cause cannot be determined.All available information has been placed on file in quality assurance for appropriate tracking, trending and follow-up.
|
| |
|
Event or Problem Description
|
|
This report is being submitted as follow up no.1 for mfg.Report no.9681834-2016-00004 to provide the retention sample evaluation results.
|
| |
|
Additional Manufacturer Narrative
|
|
As stated, this report is being submitted as follow-up no.2 for mfg.Report no.9681834-2016-00004 to provide the return sample evaluation results.The actual device has been returned to the manufacturing facility for evaluation.Visual inspection upon receipt confirmed the customer's observation.The sampling line tube had been fractured at the joint with the oxygenator outlet port.Magnifying inspection of the fracture cross-sections of the tube revealed no defects.The tube was cut vertically and subjected to dimensional testing and confirmed to meet manufacturer specifications.Electron microscopic inspection of the fracture cross-sections found that some segments were in the smooth state and other segments in the rough state and revealed no defects which would have contributed to the generation of the fracture.Functional testing was conducted and was able to reproduce the reported defect.This test was conducted on a sampling line tube of a current product sample and was exposed to a shock force at the joint of the tube and the oxygenator outlet port after having been exposed to a low temperature of 4 degrees c for 12 hours.The sampling line tube became fractured at the joint.Electron microscopic inspection of the fracture cross-section found that the state of the surface was similar to the cracked cross-section of the actual sample.There is no evidence that this event was related to a device defect or malfunction and the exact cause cannot be determined.Based on the investigation results, it is likely the actual sample was subject to a shock force in the state of being cooled by having been left under a low temperature resulting in the reported breakage of the tube, however, the definite cause of the complaint could not be determined.All available information has been placed on file in quality assurance at the manufacturing facility for appropriate tracking, trending and follow-up.
|
| |
|
Event or Problem Description
|
|
This report is being submitted as follow-up no.2 for mfg.Report no.9681834-2016-00004 to provide the return sample evaluation results.
|
| |
|
Search Alerts/Recalls
|